KFORCE Sens® electrogoniometer in the evaluation of wrist range of motion in patients with carpal tunnel syndrome: Validity and reliability

dc.contributor.authorCeylan, İsmail
dc.contributor.authorKuzu, Şafak
dc.contributor.authorCanli, Mehmet
dc.contributor.authorÖzüdoğru, Anıl
dc.contributor.authorAlkan, Halil
dc.contributor.authorTuncay, Figen
dc.date.accessioned2025-03-15T15:01:57Z
dc.date.available2025-03-15T15:01:57Z
dc.date.issued2024
dc.departmentMuş Alparslan Üniversitesien_US
dc.description.abstractThis study aimed to test the validity and reliability of the KFORCE Sens® (electrogoniometer) in carpal tunnel syndrome (CTS) patients. Thirty-one volunteers (aged 32-60 years) diagnosed with unilateral CTS were included in this study. Each of the two raters evaluated the participants' wrist range of motion (ROMs) (flexion, extension, radial deviation, and ulnar deviation) twice, for a total of four times, using an electrogoniometer. In addition, wrist ROMs were evaluated once with a universal goniometer. According to the study findings, both the inter-rater and test-retest reliability of the electrogoniometer device in all ROMs of the wrist were excellent (ICC>0.80). In addition, the ROM values of the electrogoniometer device and the ROM values of the universal goniometer in the same joint were highly correlated (r>0.80, p<0.05). The electrogoniometer is a valid, reliable, and practical device for evaluating wrist ROMs in patients with CTS. © 2024 Ondokuz Mayis Universitesi. All rights reserved.en_US
dc.identifier.doi10.52142/omujecm.41.3.10
dc.identifier.endpage509en_US
dc.identifier.issn1309-4483
dc.identifier.issue3en_US
dc.identifier.scopus2-s2.0-85213420132
dc.identifier.scopusqualityQ4
dc.identifier.startpage505en_US
dc.identifier.urihttps://doi.org/10.52142/omujecm.41.3.10
dc.identifier.urihttps://hdl.handle.net/20.500.12639/6841
dc.identifier.volume41en_US
dc.indekslendigikaynakScopusen_US
dc.indekslendigikaynakScopus
dc.language.isoen
dc.publisherOndokuz Mayis Universitesien_US
dc.relation.ispartofJournal of Experimental and Clinical Medicine (Turkey)en_US
dc.relation.publicationcategoryMakale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanıen_US
dc.rightsinfo:eu-repo/semantics/closedAccessen_US
dc.snmzKA_Scopus_20250315
dc.subjectcarpal tunnel syndromeen_US
dc.subjectelectrogoniometersen_US
dc.subjectrange of motionen_US
dc.subjectreliabilityen_US
dc.subjectwristen_US
dc.titleKFORCE Sens® electrogoniometer in the evaluation of wrist range of motion in patients with carpal tunnel syndrome: Validity and reliabilityen_US
dc.typeArticle

Dosyalar

Orijinal paket

Listeleniyor 1 - 1 / 1
Yükleniyor...
Küçük Resim
İsim:
6841.pdf
Boyut:
609.48 KB
Biçim:
Adobe Portable Document Format
Açıklama:
Tam Metin / Full Text